Type
Sandwich ELISA, Biotin-labelled antibody
Applications
Serum, Urine, Bronchoalveolar lavage, Stool, Plasma (EDTA, citrate, heparin)
Sample Requirements
Serum and plasma: 25 µl/well
For processing the stool samples we recommend using BioVendor Extraction Buffer. The reagent is not included and can be ordered separately (Cat. No.: C005821, 100 ml).
Storage/Expiration
Store the complete kit at 2–8°C. Under these conditions, the kit is stable until the expiration date (see label on the box).
Calibration Curve
Calibration Range
0.075–2.4 ng/ml
Limit of Detection
0.007 ng/ml
Intra-assay (Within-Run)
n = 8; CV = 6.6%
Inter-assay (Run-to-Run)
n = 6; CV = 7.0%
Spiking Recovery
102,70%
Dilutation Linearity
95,00%
Crossreactivity
- bovine Non-detectable
- cat Non-detectable
- dog Non-detectable
- goat Non-detectable
- hamster Non-detectable
- horse Non-detectable
- chicken Non-detectable
- monkey Non-detectable
- mouse Non-detectable
- pig Non-detectable
- rabbit Non-detectable
- rat Non-detectable
- sheep Non-detectable
- human Yes