Type
Sandwich ELISA, Biotin-labelled antibody
Applications
Serum, Cell culture supernatant
Sample Requirements
50 µl/well
Storage/Expiration
–20°C
Calibration Curve
Calibration Range
0.7–480 pg/ml
Limit of Detection
0.2 pg/ml
Intra-assay (Within-Run)
CV = 8.5%
Inter-assay (Run-to-Run)
CV = 10.4%
Spiking Recovery
117,00%
Dilutation Linearity
88,00%