Type
SandwichELISA,Biotin-labelledantibody
Applications
Serum,Urine,Plasma(EDTA,citrate,heparin)
SampleRequirements
20 µl/well
Storage/Expiration
Storethecompletekitat28°C.Undertheseconditions,thekitisstableuntiltheexpirationdate(seelabelonthe box)
CalibrationCurve
CalibrationRange
0.16 –5 ng/ml
LimitofDetection
0.04 ng/ml
Intra-assay(Within-Run)
n=8;CV=5.7%
Inter-assay(Run-to-Run)
n=7;CV=7.9%
SpikingRecovery
96.4 %
DilutationLinearity
103.6 %