Type
Sandwich ELISA, HRP-labelled antibody
Applications
Serum, Plasma-EDTA, Plasma-Heparin, Plasma-Citrate, Cell culture medium, Urine, Cerebrospinal fluid
Sample Requirements
10 µl/well
Storage/Expiration
Store the kit at 2–8°C. Under these conditions, the kit is stable until the expiration date (see label on the box).
Calibration Curve
Calibration Range
2–100 ng/ml
Limit of Detection
0.104 ng/ml
Intra-assay (Within-Run)
n = 8; CV = 2.9%
Inter-assay (Run-to-Run)
n = 6; CV = 4.7%
Spiking Recovery
108,80%
Dilutation Linearity
110,40%
Crossreactivity
- bovine Non-detectable
- cat Non-detectable
- dog Non-detectable
- goat Non-detectable
- hamster Non-detectable
- horse Non-detectable
- mouse Non-detectable
- pig Non-detectable
- rabbit Non-detectable
- rat Non-detectable
- sheep Non-detectable
- chicken Not tested
- human Yes
- monkey Yes (recommended dilution 1:5000)