Type
Sandwich ELISA, Biotin-labelled antibody
Applications
Serum
Sample Requirements
50ul/well
Storage/Expiration
Store the complete kit at 2–8°C. Under these conditions, the kit is stable until the expiration date (see label on the box).
Calibration Curve
Calibration Range
0.125–4 ng/ml
Limit of Detection
0.032 ng/ml
Intra-assay (Within-Run)
n = 8; CV = 5.4%
Inter-assay (Run-to-Run)
n = 6; CV = 7.1%
Spiking Recovery
94,50%
Dilutation Linearity
93,20%