Type
SandwichELISA,HRP-labelledantibody
Applications
Serum,Cellculturesupernatant
SampleRequirements
50 µl/well
Storage/Expiration
Storethecompletekitat2–8°C.Undertheseconditions,thekitisstableuntiltheexpirationdate(seelabelonthe box).
CalibrationCurve
CalibrationRange
6.25–400 ng/ml
LimitofDetection
1.18 ng/ml
Intra-assay(Within-Run)
CV=2.0%
Inter-assay(Run-to-Run)
CV=3.8%
SpikingRecovery
86,00%
DilutationLinearity
107,00%